QIAstat-Dx Meningitis/Encephalitis Panel Matches BioFire FilmArray in Multicenter Evaluation
A multicenter study evaluated the QIAstat-Dx ME Panel against the BioFire FilmArray ME Panel using 585 clinical and 367 contrived cerebrospinal fluid specimens. The QIAstat-Dx panel achieved 100% positive percent agreement for five bacterial and one viral target, with greater than 90% agreement for most remaining targets and greater than 99% negative percent agreement across all pathogens. The panel provides a competitive rapid diagnostic option for community-acquired central nervous system infections.
The original study
Evaluation of the QIAstat-Dx Meningitis/Encephalitis Panel, a multiplex PCR platform for the detection of community-acquired meningoencephalitis.
- Authors
- Sundelin T, Bialas J, de Diego J, Hermanowski M, Leibhan H, Ponderand L, et al.
- Journal
- Journal of clinical microbiology
- Type
- Multicenter Study, Journal Article, Research Support, Non-U.S. Gov't
- PMID
- 37702495
Original abstract
Rapid identification of the causative pathogens of central nervous system infections is essential for providing appropriate management and improving patient outcomes. The performance of QIAstat-Dx Meningitis/Encephalitis (ME) Panel-a multiplex PCR testing platform-in detecting pathogens implicated in meningitis and/or encephalitis was evaluated using BioFire FilmArray ME Panel as a comparator method. This multicenter study analyzed 585 retrospective residual cerebrospinal fluid specimens and 367 contrived specimens. The QIAstat-Dx ME Panel showed positive percent agreement (PPA) values of 100% for Neisseria meningitidis, Streptococcus agalactiae, Escherichia coli K1, Listeria monocytogenes, and Cryptococcus gattii/neoformans on clinical samples compared to the BioFire FilmArray ME Panel. The PPA values observed for Haemophilus influenzae and Streptococcus pneumoniae were 80% and 88.24%, respectively. Negative percent agreement (NPA) values were >99.0% for each of the six bacterial targets and one fungal target tested with clinical samples. One viral target, herpes simplex virus 1, exhibited a PPA value of 100.0%, while the remaining viral targets-human parechovirus, herpes simplex virus 2, human herpes virus 6, and varicella zoster virus-were >90.0%, with the exception of enterovirus, which had a PPA value of 77.8%. The QIAstat-Dx ME Panel detected five true-positive and four true-negative cases compared to BioFire FilmArray ME Panel. The NPA values for all viral pathogens were >99.0%. Overall, the QIAstat-Dx ME Panel showed comparable performance to the BioFire FilmArray ME Panel as a rapid diagnostic tool for community-acquired meningitis and encephalitis.