Lab Medicine Significance 7/10

GenMark ePlex Gram-Negative Blood Culture Panel Demonstrates High Accuracy Across 27 Targets

A multicenter clinical study assessed the GenMark Dx ePlex blood culture identification Gram-negative panel across 2,444 positive blood culture samples covering 27 organism and antimicrobial resistance gene targets. After discordant resolution, overall sample accuracy reached 92.9%, with a weighted positive percent agreement of 94.9% across all targets. The panel enables rapid molecular identification of Gram-negative pathogens and resistance determinants, potentially accelerating targeted therapy for bloodstream infections by more than 24 hours compared to conventional methods.

The original study

A Multicenter Clinical Study To Demonstrate the Diagnostic Accuracy of the GenMark Dx ePlex Blood Culture Identification Gram-Negative Panel.

Authors
Wolk DM, Young S, Whitfield NN, Reid JL, Thornberg A, Carroll KC, et al.
Journal
Journal of clinical microbiology
Type
Journal Article, Multicenter Study
PMID
34232066
Read the original study →

Original abstract

Bacteremia can progress to septic shock and death without appropriate medical intervention. Increasing evidence supports the role of molecular diagnostic panels in reducing the clinical impact of these infections through rapid identification of the infecting organism and associated antimicrobial resistance genes. We report the results of a multicenter clinical study assessing the performance of the GenMark Dx ePlex investigational-use-only blood culture identification Gram-negative panel (BCID-GN), a rapid diagnostic assay for detection of bloodstream pathogens in positive blood culture (PBC) bottles. Prospective, retrospective, and contrived samples were tested. Results from the BCID-GN were compared to standard-of-care bacterial identification methods. Antimicrobial resistance genes (ARGs) were identified using PCR and sequence analysis. The final BCID-GN analysis included 2,444 PBC samples, of which 926 were clinical samples with negative Gram stain results. Of these, 109 samples had false-negative and/or -positive results, resulting in an overall sample accuracy of 88.2% (817/926). After discordant resolution, overall sample accuracy increased to 92.9% (860/926). Pre- and postdiscordant resolution sample accuracy excludes 37 Gram-negative organisms representing 20 uncommon genera, 10 Gram-positive organisms, and 1 Candida species present in 5% of samples that are not targeted by the BCID-GN. The overall weighted positive percent agreement (PPA), which averages the individual PPAs from the 27 targets (Gram-negative and ARG), was 94.9%. The limit of detection ranged from 104 to 107 CFU/ml, except for one strain of Fusobacterium necrophorum at 108 CFU/ml.