Regulatory Significance 6/10

HER2 FISH Proficiency Testing in China Reveals Persistent Standardisation Gaps

A multicenter ring study distributing ten breast cancer samples to 169 Chinese laboratories found substantial overall agreement for HER2 FISH (Fleiss' kappa 0.765-0.911), but significantly lower concordance for cases near diagnostic cutoffs or with genetic heterogeneity (kappa 0.582). Laboratories that lacked validation after protocol updates or had no standard interpretation procedures performed significantly worse, pointing to actionable quality-improvement targets.

The original study

Interlaboratory variability of HER2 fluorescence in situ hybridization testing in breast cancer: results of a multicenter proficiency-testing ring study in China.

Authors
Peng R, Zhang K, Lin G, Li J
Journal
Diagnostic pathology
Type
Journal Article, Multicenter Study
PMID
39707446
Read the original study →

Original abstract

BACKGROUND: Accurate detection of human epidermal growth factor receptor 2 (HER2) gene amplification via fluorescence in situ hybridization (FISH) is necessary to determine HER2 status. Although many attempts have been made to increase the consistency of the results, the actual situation still needs to be determined. To investigate the latest interlaboratory variability of HER2 FISH testing for breast cancer, a multicenter proficiency-testing ring study was conducted in China. METHODS: A total of ten samples, each exhibiting distinct HER2 signal patterns and genetic heterogeneity, were distributed to 169 laboratories for HER2 FISH analysis. Data comprising both the results of the tests and feedback from questionnaires were compiled for comprehensive evaluation. RESULTS: The overall agreement among the participating laboratories was substantial to almost perfect, with a Fleiss' kappa value of 0.765-0.911. However, it is important to note that cases with characteristics of HER2 signals near the critical cutoff range or with genetic heterogeneity showed lower congruence, poorer reproducibility, and higher variability (Fleiss' kappa: 0.582). Our questionnaire showed that 52.2% (86/168) of the participants did not perform validation after their operation procedures or interpretation criteria were updated, and 75.6% (121/160) of the participants did not establish standard interpretation procedures. Since these laboratories showed worse performance (P < 0.05), the lack of validation and interpretation procedures was speculated to be the possible underlying cause. CONCLUSIONS: This study presents the latest landscape of interlaboratory variability and accuracy of HER2 FISH testing in China and highlights potential causes for the variability. Despite many years of effort, the standardization of HER2 status determination still has a long way to go.